Agilent Technologies Signal Analyzer N9030a User Manual

Page 163

Advertising
background image

3  Programming the Analyzer

STATus Subsystem

"Questionable Integrity Signal Event Query" on page 164

"Questionable Integrity Signal Negative Transition" on page 164

"Questionable Integrity Signal Positive Transition" on page 164

Questionable Integrity Signal Condition

This query returns the decimal value of the sum of the bits in the Questionable Integrity Signal
Condition register.

The data in this register is continuously updated and reflects the current conditions.

Mode

All

Remote Command

:STATus:QUEStionable:INTegrity:SIGNal:CONDition?

Example

STAT:QUES:INT:SIGN:COND?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Integrity Signal Enable

This command determines which bits in the Questionable Integrity Signal Condition Register will
set bits in the Questionable Integrity Signal Event register, which also sets the Integrity Summary
bit (bit 9) in the Questionable Register. The variable <integer> is the sum of the decimal values of
the bits you want to enable.

Mode

All

Remote
Command

:STATus:QUEStionable:INTegrity:SIGNal:ENABle <integer>

:STATus:QUEStionable:INTegrity:SIGNal:ENABle?

Example

STAT:QUES:INT:SIGN:ENAB 4

Burst Not Found will be reported to the Integrity Summary of the Status Questionable register.

Preset

32767

Min

0

Max

32767

Status
Bits/OPC
dependencies

Sequential command

Initial S/W
Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference

163