External 1 – Agilent Technologies Signal Analyzer N9030a User Manual

Page 1407

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trigger

Default Unit

S

Backwards Compatibility
Notes

The legacy :TRIGger[:SEQuence]:DELay command affects the delay for the VID, LINE, EXT1,
EXT2, and RFB triggers.

The legacy :TRIGger[:SEQuence]:OFFSet command is supported for the VIDeo, LINE, EXT1, and
EXT2 triggers.

Initial S/W Revision

Prior to A.02.00

External 1

Pressing this key, when it is not selected, selects an external input signal as the trigger. A new
sweep/measurement will start when the external trigger condition is met using the external 1
input connector on the rear panel.

Pressing this key, when it is already selected, accesses the external 1 trigger setup menu.

Key Path

Trigger

Example

TRIG:SOUR EXT1                 Swept SA measurement

TRIG:<meas>:SOUR EXT1   Measurements other than Swept  SA

Dependencies

Grayed out if in use by Point Trigger in the Source Setup menu. 

Forced to Free Run if already selected and Point Trigger is set to External 1.

State Saved

Saved in instrument state

Status Bits/OPC
dependencies

The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up
message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2
sec. This message goes away when a trigger signal appears.

Initial S/W Revision

Prior to A.02.00

Trigger Level

Sets the value where the external 1 trigger input will trigger a new sweep/measurement.

Key Path

Trigger, External 1

Remote Command

:TRIGger[:SEQuence]:EXTernal1:LEVel <level>

:TRIGger[:SEQuence]:EXTernal1:LEVel?

Remote Language Compatibility Measurement Application Reference