6  RLC Swept SA Measurement Front-Panel & SCPI Reference

System

A successful Align Now, RF will set the Last Align RF temperature to the current temperature, and
reset the Last Align RF time. A successful Align Now, All or Align Now, All but RF will set the Last
Align Now All temperature to the current temperature, and reset the Last Align Now All time. A
successful Align Now, All will also reset the Last Align RF items if the RF portion of the Align Now
succeeded.

Key Path

System, Alignments

Mode

All

Notes

The values displayed on the screen are only updated upon entry to the screen and not updated
while the screen is being displayed.

Initial S/W Revision

Prior to A.02.00

Key Path

Visual annotation in the Show Alignment Statistics screen

Mode

All

Remote Command

:SYSTem:PON:TIME?

Example

:SYST:PON:TIME?

Notes

Value is the time since the most recent start-up in seconds.

State Saved

No

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference