6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trigger

Compatibility SCPI

Initial S/W Revision

Prior to A.02.00

Trig Slope

Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.

Key Path

Trigger, External 2

Remote Command

:TRIGger[:SEQuence]:EXTernal2:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:EXTernal2:SLOPe?

Example

TRIG:EXT2:SLOP NEG

Couplings

This same slope is used in the Ext2 selection for the trigger source in the Trigger menu and for
the period timer sync source (in the Trigger menu and in the Gate Source menu).

Preset

POSitive

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:FRAMe:EXTernal2:SLOPe

Backwards Compatibility
Notes

The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.

Initial S/W Revision

Prior to A.02.00

Trig Delay

Controls a time delay during which the analyzer will wait to begin a sweep after meeting the
trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT,
but not in swept spans.

Key Path

Trigger, External 2

Remote Command

:TRIGger[:SEQuence]:EXTernal2:DELay <time>

:TRIGger[:SEQuence]:EXTernal2:DELay?

:TRIGger[:SEQuence]:EXTernal2:DELay:STATe OFF|ON|0|1

:TRIGger[:SEQuence]:EXTernal2:DELay:STATe?

Example

TRIG:EXT2:DEL:STAT ON

TRIG:EXT2:DEL 100 ms

Remote Language Compatibility Measurement Application Reference