Continuous peak search – Agilent Technologies Signal Analyzer N9030a User Manual

Page 907

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Peak Search

Above Display Line

Sets the peak table to display only the 20 highest peaks above the display line in the order
specified by the current Sort setting. If the Peak Criteria are turned on, then only peaks that meet
the defined criteria will be found. If the display line is not already on, it is turned on (it has to be
on or it cannot be used to exclude peaks).

Key Path

Peak Search, Peak Table, Peak Readout

Example

CALC:MARK:PEAK:TABL:READ GTDL

Notes

Auto return after pressed

Dependencies

When Above Display Line is selected, Display Line is turned on and becomes the active function.

Readback

Above DL

Initial S/W Revision

Prior to A.02.00

Below Display Line

Sets the peak table to display only the 20 highest peaks below the display line as defined by the
peak in the order specified by the current Sort setting. If the Peak Criteria are turned on, then
only peaks that meet the defined criteria will be found. If the display line is not already on, it is
turned on (it has to be on or it cannot be used to exclude peaks).

Key Path

Peak Search, Peak Table, Peak Readout

Example

CALC:MARK:PEAK:TABL:READ LTDL

Notes

Auto return after pressed

Dependencies

When Below Display Line is selected, Display Line is turned on and becomes the active function.

Readback

Below DL

Initial S/W Revision

Prior to A.02.00

Continuous Peak Search

Turns Continuous Peak Search on or off. When Continuous Peak Search is on, a peak search is
automatically performed for the selected marker after each sweep. The rules for finding the peak
are exactly the same as for Peak Search, including the use of the peak criteria rules. If no valid
peak is found, a “No peak found” message is generated after each sweep.

See

"More Information" on page 908

.

Remote Language Compatibility Measurement Application Reference

907