Agilent Technologies Signal Analyzer N9030a User Manual

Page 1421

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trigger

Remote Command

:TRIGger[:SEQuence]:FRAMe:ADJust <time>

Example

TRIG:FRAM:ADJ 1.2 ms

Notes

Note also that Offset is used only when the sync source is set to OFF, otherwise delay is used
(see section

"Trig Delay" on page 1429

An increase in the "offset" parameter, either from the knob or the SCPI adjust command, serves
to delay the timing of the trigger event.

Notes

The front panel interface (for example, the knob) and the :TRIG:FRAM:OFFS command adjust the
accumulated offset, which is shown on the active function display. However, the actual amount
sent to the hardware is the delta value, that is, the current offset value minus the previous offset
value.

When the SCPI command is sent, the value shown on the key (and the Active Function, if this
happens to be the active function) is updated by increasing it (or decreasing it if the value sent is
negative) by the amount specified in the SCPI command.

This is a "command only" SCPI command, with no query.

Dependencies

The invalid data indicator turns on when the offset is changed, until the next
sweep/measurement completes.

Couplings

The same offset is used in the Gate Source selection of the period timer.

Preset

0 s

State Saved

Saved in instrument state

Min

–10.000 s

Max

10.000 s

Default Unit

S

Initial S/W Revision

Prior to A.02.00

Reset Offset Display

Resets the value of the periodic trigger offset display setting to 0.0 seconds. The current displayed
trigger location may include an offset value defined with the Offset key. Pressing this key redefines
the currently displayed trigger location as the new trigger point that is 0.0 s offset. The Offset key
can then be used to add offset relative to this new timing.

Key Path

Trigger, Periodic Timer

Remote Command

:TRIGger[:SEQuence]:FRAMe:OFFSet:DISPlay:RESet

Example

TRIG:FRAM:OFFS:DISP:RES

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference

1421