5  Legacy Command Descriptions
KS, (Mixer Level)

KS, (Mixer Level)

Syntax

Legacy Products

8566A/B, 8568A/B

Description

Specifies the maximum signal level that is applied to the input mixer for a signal that is equal to
or below the reference level.

The effective mixer level is equal to the reference level minus the input attenuator setting. When
KS, is activated, the effective mixer level can be set from –10 dBm to –70 dBm in 10 dB steps.

As the reference level is changed, the coupled input attenuator automatically changes to limit the
maximum signal at the mixer input to your specified setting for signals less than or equal to the
reference level.

Format

KS, <real>DM|MV|UV

KS, OA

KS,?

Query Data Type

<real>

SCPI Equivalent Commands

[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <real> dBm

:[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?

(See

"Max Mixer Level" on page 528

)

Preset

– 10 dBm

Notes

The functions of KS, are identical to

"ML (Mixer Level)" on page 411

.

If the external amplifier gain has been set, executing IP does not reset this value. This is
to protect the instrument.