6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trace/Detector

All Traces Auto

This is designed to quickly return the selected set of detectors to the “preset” state, which is auto-
selected.

Key Path

Trace/Detector, Detector, Preset Detectors

Couplings

Sets all traces’ Detector Auto to true.

Initial S/W Revision

Prior to A.02.00

Peak / Average / NPeak

This is a setting for making a measurement of the average power and the signal envelope.

Key Path

Trace/Detector, Detector, Preset Detectors

Couplings

Trace 1: Set to peak detection, and Clear-Write.

Trace 2: Set to average detection, and Clear-Write.

Trace 3: Set to negative peak detection, and Clear-Write.

Initial S/W Revision

Prior to A.02.00

Peak / Sample / NPeak

This is a setting for making a measurement that displays a power sample and the signal envelope.

Key Path

Trace/Detector, Detector, Preset Detectors

Couplings

Trace 1: Set to peak detection, and Clear-Write.

Trace 2: Set to sample detection, and Clear-Write.

Trace 3: Set to negative peak detection, and Clear-Write.

Initial S/W Revision

Prior to A.02.00

Clear Trace

Clears the selected trace (from the front panel) or the specified trace (from SCPI). Does not affect
the state of any function or variable in the instrument. Loads mintracevalue into all of the points
in the selected trace, unless the trace is in Min Hold in which case it loads maxtracevalue. It does
this even if Update=Off.

Remote Language Compatibility Measurement Application Reference