5  Legacy Command Descriptions
CLRW (Clear Write)

The following table describes the parameters set by N9061A in the X-series instrument when the
legacy commands for trace mode and averaging are sent.

Legacy Products

X-series

Trace Commands

Averaging (VAVG)

Detector
(DET)

Trace/Detector
Trace Type

View/Blank

Detector

CLRW

Off

Normal

ClearWrite

On

Last set

CLRW

On

Sample

Trace Average

On

Sample

MXMH

Off

Peak

Max Hold

On

Peak

MXMH

On

Sample

Trace Average

On

Peak

MINH

Off

NegPeak

Min Hold

On

NegPeak

MINH

On

Sample

Trace Average

On

NegPeak

VIEW

Off

Normal

No change

View

No change

VIEW

On

Sample

Trace Average

View

Sample

BLANK

Off

Normal

No change

Blank

No change

BLANK

On

Sample

Trace Average

Blank

Sample

Legacy Products command mapping to X-series for trace/detector settings

For example, if an 8560 series analyzer receives CLRW, and averaging is set to ON, then the
analyzer's detector is automatically set to Sample. In the same circumstances, N9061A sets the X-
series instrument trace type to Trace Average, View/Blank to On, and the Detector to Sample.

Format

CLRW TRA|TRB

Preset: CLRW TRA

Query Data Type

N/A

SCPI Equivalent Commands

:TRACe1|2|3|4|5|6:TYPE WRITe (see

"Trace/Detector" on page 1321

)

Preset

TRA (after a preset, only trace A is set to clearwrite)

Notes

The functions of CLRW are identical to the command

"A1 [one] (Clear Write for Trace A)"

on page 209

and

"B1 [one] (Clear Write for Trace B)" on page 242

.

264

Remote Language Compatibility Measurement Application Reference