Agilent Technologies Signal Analyzer N9030a User Manual

Page 597

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

Notes

For R&S FSQ-B71 compatibility

Preset

IQ

Initial S/W Revision

Prior to A.02.00

I+jQ

Sets the signal input to be both the I and Q channels. The I and Q channel data will be combined
as I + j * Q.

Key Path

Input/Output, I/Q, I/Q Path

Example

Set the input to be both the I and Q channels, combined as I + j * Q.

FEED:IQ:TYPE IQ

Initial S/W Revision

Prior to A.02.00

I Only

Sets the signal input to be only the I channel. The Q channel will be ignored. The data collected is
still complex. When the center frequency is 0 the imaginary part will always be zero, but for any
other center frequency both the real and imaginary parts will be significant.

Key Path

Input/Output, I/Q, I/Q Path

Example

Set the input to be only the I channel.

FEED:IQ:TYPE IONL

Initial S/W Revision

Prior to A.02.00

Q Only

Sets the signal input to be only the Q channel. The I channel will be ignored. The Q channel will
be sent to the digital receiver block as Q+j0. The receiver's output is still complex. When the center
frequency is 0 the imaginary part will always be zero, but for any other center frequency both the
real and imaginary parts will be significant. Note that since the receiver's real output is displayed
as the "I" data, when the center frequency is 0, the Q Only input appears as the "I" data.

Key Path

Input/Output, I/Q, I/Q Path

Example

Set the input to be only the Q channel.

FEED:IQ:TYPE QONL

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference

597

Advertising