Metrohm tiamo 2.4 Manual User Manual

Page 1002

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

986

■■■■■■■■

tiamo 2.4

Titrando, 855, 867
Input range

0 - 999999 s

Default value

120 s

Titrino
Input range

0 - 9999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Titrando, 855, 867
Input range

0.1 - 999999.0 s (Increment: 0.1)

Default value

2.0 s

Titrino
Input range

0.08 - 16200 s (Increment: 0.08)

Default value

2.0 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Titrando, 855, 867
Input range

-200.0 - 200.0 µA

Selection

off

Default value

off

off
No stop.

Temperature

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.

Titrando, 855, 867
Input range

-20.0 - 150 °C

Default value

25.0 °C

Titrino
Input range

-170.0 - 500 °C

Default value

25.0 °C