Metrohm tiamo 2.4 Manual User Manual

Page 549

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

533

Slope

Minimum difference between the slope before and after the break-point.
The smaller the difference, the more break-points will be found.

Input range

0.0 - 10.0

Default value

0.9

Smoothing factor

The higher the smoothing factor, the fewer endpoints will be found.

Input range

2 - 20

Default value

5

Window

A range (window) can be defined on the measured value axis, on the vol-
ume axis or on the time axis. The break-point evaluation will only be car-
ried out in the defined window. Only the first break-point in the defined
window will be recognized.

Selection

Measured value | Volume | Time | off

Default value

off

Lower limit

Value for the lower limit of the window.

Window = Measured value
Input range

-2000.0 - 2000.0 mV

Default value

-2000.0 mV

Window = Volume
Input range

0.00000 - 9999.99 mL

Default value

0.00000 mL

Window = Time
Input range

0 - 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Window = Measured value
Input range

-2000.0 - 2000.0 mV

Default value

2000.0 mV

Window = Volume
Input range

0.00000 - 9999.99 mL

Default value

9999.99 mL

Advertising