Metrohm tiamo 2.4 Manual User Manual
Page 549
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
533
Slope
Minimum difference between the slope before and after the break-point.
The smaller the difference, the more break-points will be found.
Input range
0.0 - 10.0
Default value
0.9
Smoothing factor
The higher the smoothing factor, the fewer endpoints will be found.
Input range
2 - 20
Default value
5
Window
A range (window) can be defined on the measured value axis, on the vol-
ume axis or on the time axis. The break-point evaluation will only be car-
ried out in the defined window. Only the first break-point in the defined
window will be recognized.
Selection
Measured value | Volume | Time | off
Default value
off
Lower limit
Value for the lower limit of the window.
Window = Measured value
Input range
-2000.0 - 2000.0 mV
Default value
-2000.0 mV
Window = Volume
Input range
0.00000 - 9999.99 mL
Default value
0.00000 mL
Window = Time
Input range
0 - 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.
Window = Measured value
Input range
-2000.0 - 2000.0 mV
Default value
2000.0 mV
Window = Volume
Input range
0.00000 - 9999.99 mL
Default value
9999.99 mL