2 external measured value – Metrohm tiamo 2.4 Manual User Manual

Page 950

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

934

■■■■■■■■

tiamo 2.4

Selection

empty | ppm | % | g/L | mg/L | mg/mL | mg/
100g
| mol/L | mmol/L | mL | g | mg | µg | °C |
µL | s | mS/cm

Default value

empty

Decimal places

Number of decimal places with which the calculated measured value will
be displayed.

Input range

0 - 5

Default value

2

Comment

Comment on the calculated measured value.

Entry

250 characters

Default value

'empty'

5.6.3.11.2

External measured value

Dialog window: Method

DET/MET/SET/KFT/KFC/STAT/MEAS/DOS Proper-

ties...

Additional measured values [New]/[Properties] External measured

value

With the option Additional external measured values, a maximum of
3 new measured values can be applied and saved from other existing
measuring commands (e.g. MEAS) which are running at the same time in
other tracks. These new measured values can also be displayed in curves
under the designation Extern 1…3 and used in formulas as variables
*.EX1…3.

Command

Selection of the command from which the additional measured value will
be taken over.

Selection

Selection from existing measuring commands

Response

Selection of the quantity from the command which will be adopted as
quantity of the additional external measured value.

Selection

Selection from existing measuring quantities

Unit

Display of the unit of the external measured value.

Comment

Comment on the external measured value.