5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

684

■■■■■■■■

tiamo 2.4

Input range

–20.0 - 150.0 °C

Default value

20.0 °C

Reference temperature

The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically con-
verted to this reference temperature (usually 20 °C or 25 °C).

Input range

–20.0 - 150.0 °C

Default value

20.0 °C

Selection

off

off
The conductivity at the measuring temperature is displayed.

Temperature compensation
Definition of the type how the temperature compensation is to be carried
out.

Selection

Temperature coefficient | Sample solution

Default value

Temperature coefficient

Temperature coefficient
Manual entry of a constant temperature coefficient.
Sample solution
Selection of a function for the temperature coefficient from the table
Sample solution (TC conductivity) - %1.

Temperature coefficient

The temperature compensation is calculated on the basis of the value
entered (constant).

Input range

0.00 - 9.99 %/°C

Default value

2.00 %/°C

Sample solution

The temperature compensation is carried out on the basis of temperature-
dependent coefficients and should be used for determinations where the
sample temperature is not constant. There are two possibilities:

It is possible to use a sample solution whose temperature coefficient
has been determined by a previous measurement (MEAS TC Cond).

For ground water, spring water and surface water, the temperature
coefficients according to standard ISO 7888:1985 (German version:
DIN EN 27888:1993) are stored in the system.

Selection

DIN | Table entries

Default value

DIN