5 Method

tiamo 2.4

■■■■■■■■

797

Time
The endpoint must be exceeded during a certain period of time: the
delay time. The same delay time with different smallest possible vol-
ume increments (depending on the volume of the exchange or dosing
unit) equals different switch-off points.

NOTE

The stop conditions always lead to a stop, even if the stop criterion has
not been reached or has been switched off.

Stop drift

Titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.

Input range

1 - 999 µL/min

Default value

20 µL/min

Relative stop drift

The titration is canceled when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift value at the titration start
and the relative stop drift defined here. This parameter is displayed only if
Stop criterion = rel. drift.

Titrando
Input range

1 - 999 µL/min

Default value

10 µL/min

Delay time

If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time. The selection inf. stands for infin-
ite.

Titrando, 855
Input range

0 - 999 s

Default value

10 s

Titrino
Input range

0 - 999 s

Default value

10 s

Selection

inf.