5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

886

■■■■■■■■

tiamo 2.4

Titrando, 855
Input range

0.0 - 150.0 °C

Default value

0.2 °C

Action

Selection of the event to be triggered if the upper or lower limit is viola-
ted:

Selection

Cancel determination | Cancel command
(Titrando only)
| Wait for [Continue] | Wait for
limit ok
| none

Default value

none

Cancel determination
The running STAT pH command will be canceled, then the exit track
(if present) will be started and the determination will be finished.
Cancel command (Titrando only)
The running STAT pH command will be canceled, then the next com-
mand will be carried out.
Wait for [Continue]
Reagent dosing in the running STAT pH command will be interrupted
and a message will be displayed. As soon as the monitored measured
value is again within the limits (including hysteresis), reagent dosing
can be resumed by pressing [Continue] in this message box.
Wait for limit ok
Reagent dosing in the running STAT pH command will be interrupted.
As soon as the monitored measured value is again within the limits
(including hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are violated.

Track call on limit exceeding
In this table, which cannot be edited directly, a maximum of 20 entries
can be defined as to which track is to be started automatically when a
particular limit value is violated.

[New]

Opens the Track call # dialog window, in which the parameters for the
track call can be entered.

[Properties]

Opens the Track call # dialog window, in which the parameters for the
track call can be edited (see Chapter 5.6.3.8.4, page 863).

[Delete]

Deletes the track call selected in the table.