3 cal spec - measuring parameters, 4 cal spec - calibration wavelengths – Metrohm tiamo 2.4 Manual User Manual

Page 1128

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1112

■■■■■■■■

tiamo 2.4

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.

If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.

Selection

Avantes Spectrometer

Default value

Avantes Spectrometer

5.6.5.5.3

CAL Spec - Measuring parameters

Tab: Method

CAL Spec Properties... Measuring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.

Integration time

Integration time for spectrum recording.

Input range

0.01 - 600000 ms

Default value

6 ms

Averaged spectra

Number of spectra that are recorded and averaged.

Input range

1 - 10000

Default value

10

5.6.5.5.4

CAL Spec - Calibration wavelengths

Tab: Method

CAL Spec Properties... Calibration wavelengths

Command name

Name of the command.

Entry

25 characters