3 meas cond - measuring parameters, Measuring parameters – Metrohm tiamo 2.4 Manual User Manual

Page 1038

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1022

■■■■■■■■

tiamo 2.4

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Input range

-15 - 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is activated, the stirrer will be switched off automatically
when the command has finished.

5.6.4.10.3

MEAS Cond - Measuring parameters

Tab: Method

MEAS Cond Properties... Measuring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measurement

Measurement frequency

712
Selection

auto | 300 Hz | 2.4 kHz

Default value

auto

auto
The more suitable of the two frequencies is selected.

Type of measurement

The two measuring modes Measurement with drift control or Mea-
surement without drift control
can be selected.

Selection

Measurement with drift control | Measure-
ment without drift control

Default value

Measurement with drift control

Measurement with drift control
The measurement is carried out drift-controlled if this option is
selected. The measurement is canceled as soon as the defined Signal
drift
or the Stop measured value is reached or a Waiting time has
elapsed. The corresponding parameters are not visible if this option is
disabled. This option is not available for the 712 Conductometer.

Advertising