5 Method

tiamo 2.4

■■■■■■■■

1197

Lower hysteresis

Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value
lower limit ok
is triggered.

Titrando, 855
Input range

0.0 - 2000.0 mV

Default value

2.0 mV

Upper limit

Upper limit of the measured value. If the measured value exceeds this
limit, then the event Measured value upper limit exceeded is trig-
gered.

Titrando, 855
Input range

-2000.0 - 2000.0 mV

Default value

2000.0 mV

Titrino
Input range

-2000 - 2000 mV

Default value

2000 mV

Upper hysteresis

The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.

Titrando, 855
Input range

0.0 - 2000.0 mV

Default value

2.0 mV

Action

Selection of the event to be triggered if the upper or lower limit is
exceeded:

Selection

Stop determination | Stop command (only
Titrando)
| Wait for [Continue] | Wait for limit
ok
| none

Default value

none

Stop determination
The running DOS command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running DOS command is quit, then the next command is exe-
cuted.