5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

608

■■■■■■■■

tiamo 2.4

Titrino
Input range

-170.0 - 500.0 °C

Default value

25.0 °C

5.6.3.3.3.5

MET U - Stop conditions

Tab: Method

MET U Properties... Stop conditions

Command name

Name of the command.

Entry

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.

Stop volume

Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.

Titrando, 855
Input range

0.00000 - 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Input range

0.000 - 9999.99 mL

Default value

100.00 mL

Selection

off

Stop measured value

Stops when the preset value for a measuring point has been exceeded or
not achieved since the start of the titration.

Titrando, 855
Input range

-2000.0 - 2000.0 mV

Selection

off

Default value

off

Titrino
Input range

-2000 - 2000 mV

Selection

off

Default value

off