Metrohm tiamo 2.4 Manual User Manual

Page 1402

Advertising
background image

6.6 Subwindow Titrants/Solutions

■■■■■■■■■■■■■■■■■■■■■■

1386

■■■■■■■■

tiamo 2.4

Cylinder serial number

Serial number of the cylinder. It is printed on new cylinders. With intelli-
gent exchange units it is read out automatically. It can be modified at any
time, e.g. if the cylinder has to be replaced.

IEU
Entry

8 characters

Default value

'empty'

EU
Entry

10 characters

Default value

'empty'

Parameters for preparation
Configuration of the parameters which are to be used in the PREP com-
mand.

Volume

Volume to be dosed in during preparation.

Input range

0.00000 - 99,999.99999 mL

Selection

Cylinder volume

Default value

Cylinder volume

Cylinder volume
The whole contents of the cylinder will be dosed.

Cycles

Number of rinsing cycles during preparation.

Input range

1 - 9

Default value

2

Dosing rate

Rate at which the solution is to be dosed. The maximum dosing rate
depends on the cylinder volume of the exchange unit used (see below).
When the function is carried out, the rate will automatically be reduced to
the highest possible value.

Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum

Filling rate

Rate at which the cylinder is to be filled. The maximum filling rate depends
on the cylinder volume of the exchange unit used (see below). When the

Advertising