5 Method

tiamo 2.4

■■■■■■■■

901

5.6.3.8.7.4

STAT U - Control parameters

Tab: Method

STAT U Properties... Control parameters

Command name

Name of the command.

Entry

25 characters

Control point

Control point at

Definition of the voltage value to be controlled.

Titrando, 855
Input range

–2,000.0 - 2,000.0 mV

Selection

off

Default value

off

Titrino
Input range

–2,000 - 2,000 mV

Selection

off

Default value

off

Titration rate

Titration rate

Three predefined rates, for which the parameters for Control are not dis-
played, can be selected for the titration rate. The Users setting must be
selected in order to be able to edit these parameters.

Selection

50 µL/min | 100 µL/min | 500 µL/min | Users

Default value

100 µL/min

Control

NOTE

Will be displayed only if the Users option is selected in the selection list
Titration rate.

Dynamics

The control range defines the measured value range before the specified
control point. It has a decisive influence on the titration rate and thus also
on the accuracy (see Chapter 5.6.3.8.1, page 860).