4 stat u - control parameters, Control parameters – Metrohm tiamo 2.4 Manual User Manual
Page 917
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
901
5.6.3.8.7.4
STAT U - Control parameters
Tab: Method
▶ STAT U ▶ Properties... ▶ Control parameters
Command name
Name of the command.
Entry
25 characters
Control point
Control point at
Definition of the voltage value to be controlled.
Titrando, 855
Input range
–2,000.0 - 2,000.0 mV
Selection
off
Default value
off
Titrino
Input range
–2,000 - 2,000 mV
Selection
off
Default value
off
Titration rate
Titration rate
Three predefined rates, for which the parameters for Control are not dis-
played, can be selected for the titration rate. The Users setting must be
selected in order to be able to edit these parameters.
Selection
50 µL/min | 100 µL/min | 500 µL/min | Users
Default value
100 µL/min
Control
NOTE
Will be displayed only if the Users option is selected in the selection list
Titration rate.
Dynamics
The control range defines the measured value range before the specified
control point. It has a decisive influence on the titration rate and thus also
on the accuracy (see Chapter 5.6.3.8.1, page 860).