5 Method

tiamo 2.4

■■■■■■■■

547

Initial measured value

NOTE

Only displayed for Titrando and 855.

The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:

Measured value drift

Drift for the measured value acceptance during the titration.

Titrando, 855
Input range

0.01 - 99.90 µA/min

Default value

20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Titrino
Input range

0.05 - 99.90 µA/min

Default value

20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.

Input range

0 - 999999 s

Default value

0 s

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.