4 stdadd auto - measuring parameters – Metrohm tiamo 2.4 Manual User Manual

Page 1109

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

1093

Selection

mol/L | mmol/L | µmol/L | ppm | g/L | mg/L |
µg/L | mg/mL | µg/mL | % | mEq/L

Default value

ppm

Control parameters

Dosing rate

Speed at which the volume increments are added.

Selection

slow | medium | fast

Default value

fast

Delta U

Potential difference that is to be achieved by a standard addition.

Input range

1 - 999 mV

Default value

10 mV

5.6.4.19.4

STDADD auto - Measuring parameters

Tab: Method

STDADD Properties... Measuring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measurement with drift control
The measurement is stopped as soon as the defined Signal drift or the
Stop measured value is reached or a defined Waiting time has
elapsed.

Measured value drift

The measured value is only accepted if the signal drift defined here has
been fallen short of.

Input range

0.1 - 999.0 mV/min

Default value

10.0 mV/min

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the waiting time is elapsing.

Input range

0 - 999999 s

Default value

0 s