5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

820

■■■■■■■■

tiamo 2.4

auto
The titration direction is determined automatically from the initial mea-
sured value and the set endpoint.

Extraction time

The titration is not canceled until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, for instance, for the titration of sparingly soluble sam-
ples.

Input range

0 - 999,999 s

Temperature

Manually entered titration temperature. If a temperature sensor is con-
nected and the Temperature measurement is set on the General/
Hardware
tab under Sensor to automatic or continuous, then the
temperature will be measured continuously. This value is used for temper-
ature correction in pH measurements.

Titrando, 855
Input range

-20.0 - 150.0 °C

Default value

25.0 °C

Titrino
Input range

-170.0 - 500.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando
Input range

0.1 - 999,999 s

Default value

2.0 s

Titrino
Input range

0.08 - 16,200 s

Default value

2.00 s

5.6.3.5.3.6

KFT Upol - Stop conditions

Tab: Method

KFT Upol Properties... Stop conditions

Command name

Name of the command.

Entry

25 characters

Conditions for stopping the titration. If several stop conditions are set,
then the criterion which is fulfilled first will stop the titration.