5 Method

tiamo 2.4

■■■■■■■■

1209

Titrando, 846, 856, 867
Selection

1 | 2 | 3 | 4

Default value

1

778, 789, 814, 815, 855, 864, 874
Selection

1 | 2 | 3

Default value

1

730, 774
Selection

1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12

Default value

1

Titrino
Selection

internal D0

736, 751, 758, 799
Selection

internal D0 | external D1 | external D2

Default value

internal D0

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked. At the start of the
command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.

Entry

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

5.6.6.6

EMPTY

5.6.6.6.1

EMPTY - Overview

Dialog window: Method

EMPTY Properties... EMPTY - 'Command name'

Command for Emptying cylinder and tubing of a dosing unit.

Devices
This command can be executed with the following devices:

Titrando: 808, 809, 835, 836, 841, 842, 851, 852, 857, 859, 888, 890,
901, 902, 904, 905, 906, 907