Metrohm tiamo 2.4 Manual User Manual

Page 1021

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

1005

Identifica-
tion

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has not ever been started

.DBL

Total duration for the processing of the command in s

.EGF

Last measured gas flow rate (measured value following
processing of the command) in mL/min

.EME

Final measured value (measured value after processing
of the command) in the unit of the measured value

.FIN

Command status; 1 = Command has been ended
at least once
, 0, invalid (variable not available) =
Command has never been ended

.GMA

Maximum gas flow rate in mL/min

.GMI

Minimum gas flow rate in mL/min

.GMN

Average gas flow rate in mL/min

.IGF

Initial gas flow (measured value at the time of the start
of the command) in mL/min

.IME

Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value

.LP.CAx

Calculated value × (1...3) for the last measuring point
of the measuring point list

.LP.EXx

External value × (1...3) for the last measuring point of
the measuring point list

.LP.GFL

Gas flow value for the last measuring point of the
measuring point list in mL/min

.LP.MEA

Measured value (temperature) for the last measuring
point of the measuring point list in the unit of the
measured value

.LP.TEM

Temperature for the last measuring point of the mea-
suring point list in °C

.LP.TIM

Time in s until the last measuring point of the measur-
ing point list is reached

.MA.MEA

Maximum measured value (temperature) in the unit of
the measured value