Metrohm tiamo 2.4 Manual User Manual

Page 929

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

913

The following methods for the evaluation of titration curves can be acti-
vated and defined on this tab:

Rate evaluation

Rate evaluation

on | off (Default value: off)

If this check box is activated, a maximum of nine time slots can be
defined, in which the mean dosing rate can be determined by linear
regression. The defined time slots are shown in the window table and can
be edited with the following buttons:

[New]

Opens the Evaluation window # dialog window, in which the parame-
ters for a new time slot can be entered (see Chapter 5.6.3.8.3, page 862).

[Properties]

Opens the Evaluation window # dialog window, in which the parame-
ters for the selected time slot can be edited (see Chapter 5.6.3.8.3, page
862)
.

[Delete]

Deletes the time slot selected in the table.

Fixed endpoint evaluation

Fixed endpoint evaluation

on | off (Default value: off)

If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value, Time or Volume) for the fixed endpoint from the
measuring point list.

[New]

Opens the Fixed endpoint evaluation # dialog window for entering a
new fixed endpoint (see Chapter 5.6.3.8.7.10, page 916).

[Properties]

Opens the Fixed endpoint evaluation # dialog window to edit the
fixed endpoint selected in the table (see Chapter 5.6.3.8.7.10, page
916)
.

[Delete]

Deletes the selected line.

Advertising