Metrohm tiamo 2.4 Manual User Manual

Page 856

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

840

■■■■■■■■

tiamo 2.4

Selection

automatic | manual | off

Default value

automatic

automatic
The drift lastly determined during conditioning will be used as drift
value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.

Drift value

Drift value for drift correction. This parameter is only shown for Drift cor-
rection
= manual.

Input range

0.0 - 99.9 µg/min

Default value

0.0 µg/min

Stop time

Determination stops when the specified time has elapsed since the start of
the conditioning.

Input range

1 - 999999 s

Selection

off

Default value

off

Stabilizing time

Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.

Input range

0 - 999 s

Default value

0 s

Automatic start after sample addition time

on | off (Default value: on)

If this option is activated, then the titration will be started automatically
after the sample addition time.

Sample addition time

Waiting time between stop of conditioning and start of titration during
which the sample can be added.

Input range

0 - 99 s

Default value

0 s