5 Method

tiamo 2.4

■■■■■■■■

723

not defined
No tests will be carried out.

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando (without 888)
Selection

1 | 2

Default value

1

855, 888
Selection

1

Default value

1

Titrino
Selection

1 | 2 | diff.

Default value

1

Sensor

Selection of a sensor of type pH electrode, metal electrode, pH elec-
trode
or ISE electrode from the list of sensors defined in the sensor
table. The calibration data for the sensor will be adopted for pH and ISE
electrodes.

Selection

Sensor name | pH electrode | Metal electrode |
ISE electrode

Default value

Metal electrode

Temperature measurement

Type of temperature measurement.

Titrando, 855
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will used.