Metrohm tiamo 2.4 Manual User Manual

Page 1200

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1184

■■■■■■■■

tiamo 2.4

Titrando, 855
Input range

0.0 - 150.0 °C

Default value

0.2 °C

Action

Selection of the event to be triggered if the upper or lower limit is
exceeded:

Selection

Stop determination | Stop command (only
Titrando)
| Wait for [Continue] | Wait for limit
ok
| none

Default value

none

Stop determination
The running DOS command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running DOS command is quit, then the next command is exe-
cuted.
Wait for [Continue]
Reagent dosing in the current DOS command is interrupted and a
message appears. As soon as the monitored temperature is again
within the limits (including hysteresis) reagent dosing can be resumed
by pressing on [Continue] in this message box.
Wait for limit ok
Reagent dosing in the current DOS command is interrupted. As soon
as the monitored temperature is once again within the limits (including
hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.

Track call on limit exceeding
In this table, which cannot be edited directly, max. 20 entries can be
defined for the track that is to be started automatically when a particular
limit is exceeded.

[New]

Opens the dialog window Track call in which the parameters for a new
track call can be entered (see Chapter 5.6.6.3.3, page 1170).

[Properties]

Opens the dialog window Track call in which the parameters for the
selected track call can be edited (see Chapter 5.6.6.3.3, page 1170).

[Delete]

Deletes the track call selected in the table.