4.5 Determination overview

■■■■■■■■■■■■■■■■■■■■■■

344

■■■■■■■■

tiamo 2.4

lated from these measured values. Listed below are the individual mea-
sured values and the results for Slope, E(0), c(blank) and Variance.

Calibration curve for CAL LOOP Opt
In the dialog window Calibration curve/Calibration data a tab marked
with the command name showing the calibration curve and calibration
data will be displayed for each CAL LOOP Opt command. The command
type is shown above the calibration curve. The curve display shows the
measured values and the curve calculated from these measured values.
Listed below are the individual measured values and the results: conc
['Unit']
,

ε[mAU], T [°C], t [s], Standard #, Wavelength, c0, c1, c2,

c3, Confidence interval, Number of outliers and Calibration range.

Standard addition curve for STDADD
In the dialog window Calibration curve/Calibration data a tab marked
with the command name showing the standard addition curve and evalu-
ation data will be displayed for each STDADD command (see Chapter
5.6.4.17.1, page 1072)
. The command type is shown above the calibration
curve. The curve display shows the measured values and the curve calcu-
lated from these measured values. Listed below are the individual mea-
sured values and the results for Determination, Show calibration
curve...
, E(0), Determination, Show calibration curve... and Var-
iance
.

Calibration data for CAL Cond
In the dialog window Calibration curve/Calibration data a tab marked
with the command name showing the calibration data will be displayed
for each CAL Cond command (see Chapter 5.6.5.4.1, page 1101). Dis-
played are the command type, the measured values and the cell constant
calculated with these measured values.

Calibration curve for ELT LOOP
In the dialog window Calibration curve/Calibration data a tab marked
with the command name showing the calibration curve and calibration
data will be displayed for each CAL Cond command (see Chapter
5.6.5.8, page 1122)
. The command type is shown above the calibration
curve. The curve display shows the measured values and the curve calcu-
lated from these measured values. Listed below are the individual mea-
sured values and the results for Slope, pH(0) and Variance.