5 Method

tiamo 2.4

■■■■■■■■

823

Input range

0.0 - 99.9 µL/min

Default value

0.0 µL/min

Stop volume

Stop of the determination when the specified volume has been dosed
since the start of the conditioning. The stop volume should be adjusted to
the size of the titration cell in order to prevent any overflow.

Titrando
Input range

0.000000 - 9,999.99 mL

Default value

20.00 mL

Selection

off

Default value

off

Stop time

Stop of the determination when the specified time has elapsed since the
start of the conditioning.

Titrando
Input range

1 - 999,999 s

Selection

off

Default value

off

Stabilizing time

Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.

Input range

0 - 999 s

Default value

0 s

Automatic start after sample addition time

on | off (Default value: on)

If this option is activated, then the titration will be started automatically
after the sample addition time.

Sample addition time

Waiting time between stop of conditioning and start of titration during
which the sample can be added.

Input range

0 - 99 s

Default value

0 s