Metrohm tiamo 2.4 Manual User Manual

Page 1041

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

1025

Selection

off

Default value

off

off
No stop.

Temperature

Measuring temperature

Temperature of the solution to be measured. If a temperature sensor is
connected, the temperature entered here is ignored.

712
Input range

-170.0 - 500.0 °C

Default value

20.0

856
Input range

-20.0 - 150.0 °C

Default value

20.0

Reference temperature

The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically con-
verted to this reference temperature (usually 20 °C or 25 °C).

712
Input range

-170.0 - 500.0 °C

Default value

20.0

856
Input range

-20.0 - 150.0 °C

Default value

20.0

Selection

off

off
The conductivity at the measuring temperature is displayed.

Temperature compensation
Definition of the type how the temperature compensation is to be carried
out.

Selection

Temperature coefficient | Sample solution

Default value

Temperature coefficient

Temperature coefficient
Manual entry of a constant temperature coefficient.