5 Method

tiamo 2.4

■■■■■■■■

1183

Monitoring Temperature

Monitoring Temperature

on | off (Default value: off)

If this option is switched on then the temperature will be monitored and
any exceeded limits will be entered in the measuring point list.

Lower limit

Lower limit of the temperature. When the temperature falls below this
value, the action Lower temperature limit exceeded is triggered.

Titrando, 855
Input range

-20.0 - 150.0 °C

Default value

-20.0 °C

Titrino
Input range

-170.0 - 500.0 °C

Default value

-170.0 °C

Lower hysteresis

Lower hysteresis of the temperature. When the temperature once again
exceeds the lower limit by this hysteresis value, the action Temperature
lower limit ok
is triggered.

Titrando, 855
Input range

0.0 - 150.0 °C

Default value

0.2 °C

Upper limit

Upper limit of the temperature. When the temperature exceeds this value,
the action Temperature upper limit exceeded is triggered.

Titrando, 855
Input range

-20.0 - 150.0 °C

Default value

150.0 °C

Titrino
Input range

-170.0 - 500.0 °C

Default value

500.0 °C

Upper hysteresis

Upper hysteresis of the temperature. When the temperature once again
falls below the upper limit by this hysteresis value, the action Tempera-
ture upper limit ok
is triggered