5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

630

■■■■■■■■

tiamo 2.4

Start measured value

Once the start measured value is reached, the dosing of the start volume
is stopped and the next start condition is processed or the titration is
started. If the start measured value is achieved by the addition of a start
volume, then titration will start directly.

Input range

-2000.0 - 2000.0 mV

Selection

off

Default value

off

Dosing rate

Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the
exchange unit or dosing unit used.

Input range

0.01 - 166.00 mL/min

Default value

5.00 mL/min

Selection

maximum

Start slope

NOTE

Only displayed for Titrando and 855.

Start slope

When the start slope is reached the dosing of the start volume is stopped
and the titration is started. If the start slope is achieved by the dosing of a
start volume then the titration starts directly.

Input range

0 - 999 mV/mL

Selection

off

Default value

off

Dosing rate

Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the
exchange unit or dosing unit used.

Input range

0.01 - 166.00 mL/min

Default value

5.00 mL/min

Selection

maximum