5 Method

tiamo 2.4

■■■■■■■■

1179

Input range

0 - 999999 s

Default value

0 s

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously. This value is used for
temperature correction during monitoring with pH measurements.

Titrando, 855
Input range

-20.0 - 150.0 °C

Default value

25.0 °C

Titrino
Input range

-170.0 - 500.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando, 855
Input range

0.1 - 999999 s

Default value

5.0 s

Titrino
Input range

1 - 999999 s

Default value

5 s

5.6.6.3.4.4

DOS pH - Stop conditions

Tab: Method

DOS pH Properties... Stop conditions

Command name

Name of the command.

Entry

25 characters

The conditions for stopping the dosing are defined on this tab. If more
than one stop condition is in effect at the same time, then the criterion
which is fulfilled first will stop the dosing.

Stop volume

Stops when the given volume has been added after the start of the dos-
ing. The stop volume should be adapted to suit the sample weight or the
titration vessel size.