Metrohm tiamo 2.4 Manual User Manual

Page 1246

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1230

■■■■■■■■

tiamo 2.4

Stirrer

Stirrer

on | off (Default value: off)

Selection of the stirrer.

Titrando, 730, 846, 856, 867
Selection

1 | 2 | 3 | 4

Default value

1

778, 789
Selection

T1 | T2 (tower stirrer) | 1 … 3 (MSB stirrer)

Default value

T1

814, 815, 855, 864, 874
Selection

1 | 2 | 3

Default value

1

Titrino, 756, 831
Selection

1

Stirrer type

Display of the stirrer type. Unknown is displayed either with devices
unable to read out the stirrer type or if no stirrer is connected

Selection

801 Magnetic Stirrer | 802 Rod Stirrer | 803 Ti
Stand magnetic stirrer
| 804 Ti Stand rod stir-
rer
| unknown

Stirring rate

Setting the stirring rate or rotational speed. The algebraic sign of the stir-
ring rate changes the direction in which the stirring is done. The default
setting 8 corresponds to a rotational speed of 1000 rpm.

Titrando, 814, 815, 846, 855, 856, 867, 864, 874
Input range

-15 - 15

Default value

8

730, 778, 789
Input range

0 - 15

Default value

8