5 signatures, Electronic signature, Signatures – Metrohm tiamo 2.4 Manual User Manual

Page 1335

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.4

■■■■■■■■

1319

Selection

on | off

Default value

off
FDA default: on (can no longer be edited)

Comment on modification of determinations

If this check box is activated, then each time that a determination is
modified a modification reason and a modification comment must be
entered that will then be saved in the determination and displayed in the
database in the Information subwindow on the Determination tab.
The reason and comments are also recorded in the Audit Trail.

Selection

on | off

Default value

off
FDA default: on (can no longer be edited)

Comment on modification of sample data (live)

If this check box is activated then each time that a determination is
modified a modification reason and a modification comment must be
entered; these are saved in the determination and shown in the database
in the Information subwindow on the Sample tab. The reason and
comments are also recorded in the Audit Trail.

Selection

on | off

Default value

off
FDA default: on (can no longer be edited)

6.2.2.5

Signatures

Tab: Configuration

Tools Security settings Signatures

The parameters for the electronic signature can be configured on the Sig-
natures
tab.

Inactivity delay

If this check box is activated, then the dialog window for signing will be
automatically closed when the entered time limit has expired.

Selection

on | off

Default value

off

Input range

1 - 60 min

Default value

10 min

Remove password after signature

If this check box is activated, then the password must be entered again
after each signature.

Selection

on | off

Default value

off
FDA default: on (can no longer be edited)