5 Method

tiamo 2.4

■■■■■■■■

909

Monitoring Dosing rate

Monitoring Dosing rate

on | off (Default value: off)

If this option is enabled, then the averaged dosing rate will be monitored
and any violated limits will be entered in the measuring point list. The
monitoring of the dosing rate is not started until 10 s after the start of the
titration.

Lower limit

Lower limit of the dosing rate. If the dosing rate falls below this limit, the
event Dosing rate lower limit violated is triggered.

Titrando, 855
Input range

0.00 - 166.00 mL/min

Default value

0.00 mL/min

Titrino
Input range

0.00 - 150.00 mL/min

Default value

0.00 mL/min

Lower hysteresis

Lower hysteresis of the dosing rate. If the dosing rate exceeds the lower
limit by this hysteresis value again, then the event Dosing rate lower
limit OK
is triggered.

Titrando, 855
Input range

0.00 - 166.00 mL/min

Default value

0.20 mL/min

Upper limit

Upper limit of the dosing rate. If the dosing rate exceeds this limit, then
the event Dosing rate upper limit violated is triggered.

Titrando, 855
Input range

0.00 - 166.00 mL/min

Default value

166.00 mL/min

Titrino
Input range

0.00 - 150.00 mL/min

Default value

150.00 mL/min