5 Method

tiamo 2.4

■■■■■■■■

1091

Selection

'Sensor name' | ISE electrode

Default value

ISE electrode

Temperature measurement

Type of temperature measurement.

Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
under Temperature on the Measuring parameters tab will be
used.
off
The temperature will not be measured. The temperature manually
entered under Temperature on the Measuring parameters tab will
be used.

Stirrer

Stirrer

Selection of the stirrer.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Input range

-15 - 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically
when the command has finished.