5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

796

■■■■■■■■

tiamo 2.4

Titrando
Input range

0.01 - 166.00 mL/min

Default value

1.00 (slow), maximum (optimal), maximum
(fast) mL/min

Selection

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Default value

1.00 (slow), maximum (optimal), maximum
(fast) mL/min

Selection

maximum

Min. volume increment

This parameter determines the addition rate with which dosing is carried
out at the very beginning and at the end of the titration. The smaller the
increment that is selected, the slower the titration and the higher the
accuracy.

Input range

0.1 - 9.9 µL

Default value

minimum (slow), minimum (optimal), 5.0
(fast) µL

Stop criterion

Stop criterion

The titration is canceled after the last dosing once the endpoint has been
reached and the stop criterion has been fulfilled. The titration can be can-
celed when a particular Drift has been reached or after a preset Time.

Titrino
Selection

Drift | Time

Default value

Drift

Titrando
Selection

Drift | rel. drift | off | Time

Default value

Drift

Drift
The titration is canceled when the Stop drift has been reached.
rel. drift
The titration is canceled when the relative stop drift (drift value at the
start + Relative stop drift) has been reached.
off
The titration is not canceled.