5 Method

tiamo 2.4

■■■■■■■■

609

Stop EP

The titration is stopped when the specified number of equivalence points
has been found.

Input range

1 - 9

Default value

9

Selection

off

Volume after EP

When the number of equivalence points defined under Stop EP has been
found, this volume will be added. In this way you can see the curve shape
after the equivalence point is reached.

Titrando, 855
Input range

0.01000 - 9999.99 mL

Selection

off

Default value

off

Stop time

Stops when the preset time (including start conditions) has elapsed since
the start of the titration.

Titrando, 855
Input range

0 - 999999 s

Selection

off

Default value

off

Filling rate

Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.

Titrando, 855
Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum