5 Method

tiamo 2.4

■■■■■■■■

1157

Measurement with drift control
The pH calibration is carried out drift-controlled. The measurement is stop-
ped as soon as the Signal drift is reached or a defined waiting time has
elapsed.

Signal drift

The measured value is only accepted if the signal drift defined here has
been fallen short of.

Titrando, 855, 867
Input range

0.1 - 999.0 mV/min

Default value

2.0 mV/min

Titrino
Input range

0.5 - 9990. mV/min

Default value

2.0 mV/min

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the waiting time is elapsing.

Titrando, 855, 867
Input range

0 - 999999 s

Default value

10 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered then a waiting time that is suitable for
the drift will be calculated automatically according to the following equa-
tion:

Titrando, 855, 867
Input range

0 - 999999 s

Default value

110 s

Titrino
Input range

0 - 9999 s

Default value

110 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.