Metrohm tiamo 2.4 Manual User Manual

Page 761

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

745

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando (without 888)
Selection

1 | 2

Default value

1

855, 888
Selection

1

Default value

1

Sensor

Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.

Selection

Sensor name | Metal electrode

Default value

Metal electrode

I(pol)

The polarization voltage is the voltage applied to the polarizable electrode
during a amperometric measurement.

Titrando, 855
Input range

–125.0 - 125.0 µA (Increment: 0.5)

Default value

5.0 µA

Titrino
Input range

–127 - 127 µA (Increment: 1)

Default value

5 µA

Electrode check

on | off (Default value: off)

If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.

Temperature measurement

Type of temperature measurement.

Titrando, 855
Selection

continuous | automatic | off

Default value

automatic