Metrohm tiamo 2.4 Manual User Manual

Page 766

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

750

■■■■■■■■

tiamo 2.4

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.
fast
For less critical, fast titrations.
User
Editing the individual titration parameters which affect the titration
rate.

Control

NOTE

Will be displayed only if the Users option is selected in the selection list
Titration rate.

Dynamics

The control range defines the measured value range before the specified
endpoint. It has a decisive influence on the titration rate and thus also on
the accuracy. Within the control range, dosing is controlled by the Min.
rate
. The closer the endpoint, the slower the dosing until the Min. rate
has been reached. The larger the control range, the slower the titration.
Outside the control range, dosing is carried out continuously at Max.
rate
(see Chapter 5.6.3.4.2, page 695).

Titrando, 855
Input range

0.1 - 2,000.0 mV

Default value

300.0 (slow), 100.0 (optimal), 30.0 (fast) mV

Selection

off

off
The entire measured value range is defined as control range. The mini-
mum titration rate is applied throughout the entire titration.

Titrino
Input range

1.00 - 2,000 mV

Default value

300.0 (slow), 100.0 (optimal), 30.00 (fast) mV

Selection

off

off
The entire measured value range is defined as control range. The mini-
mum titration rate is applied throughout the entire titration.