7.16 RS-232 device

■■■■■■■■■■■■■■■■■■■■■■

1574

■■■■■■■■

tiamo 2.4

Timeout

Maximum waiting time for receiving characters. If this time is exceeded,
receiving will be stopped.

Input range

1000 - 9999 ms

Default value

2000 ms

Terminator for send

Terminator for send in hexadecimal code.

Selection

\0D | \0D\0A | \0D\0A\0A | \0D\0D\0A

Default value

\0D\0A
(\0D = CR, \0A = LF)

Terminator for receiving

Final character for receiving in hexadecimal code.

Selection

\0D | \0D\0A | \0D\0A\0A | \0D\0D\0A

Default value

\0D\0A
(\0D = CR, \0A = LF)

Code page

Code page used for the data transfer.

Selection

Cp437 | Cp850 | Cp852 | ASCII

Default value

Cp437

Sending characters separately
on | off (Default value: off)

If this option is enabled each character is sent separately. The delay time is
allowed to run before sending the next character.

Delay time

Delay time until the sending of the next character.
Input range

0 - 9999 ms

Default value

0 ms

[Connect]

Establishes the connection to the RS-232 device. The dialog window
Establish connection appears (see Chapter 7.16.5, page 1576).

[Disconnect]

Disconnects the connection to the RS-232 device (only then the device
can be deleted out of the configuration).