Metrohm tiamo 2.4 Manual User Manual

Page 832

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

816

■■■■■■■■

tiamo 2.4

End point

EP at

Measured value for the endpoint.

Input range

–200.0 - 200.0 µA

Default value

25.0 µA

Selection

off

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-
played. The User setting must be selected in order to be able to edit these
parameters.

Selection

slow | optimal | fast | User

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.
fast
For less critical, fast titrations.
User
Editing the individual titration parameters which affect the titration
rate.

Control

NOTE

Will be displayed only if the user option is selected in the selection list
Titration rate.

Dynamics

The control range defines the measured value range before the specified
endpoint. It has a decisive influence on the titration rate and thus also on
the accuracy. Within the control range, dosing is controlled by the Min.
rate
. The closer the endpoint, the slower the dosing until the Min. rate

Advertising