6.3 Configuration data

■■■■■■■■■■■■■■■■■■■■■■

1348

■■■■■■■■

tiamo 2.4

which the measured voltage has reached this comparison value except for
±1 mV.

Input range

0 - 999 s

Default value

60 s

Min. offset potential

Minimum permitted offset voltage (voltage at pH = 7.0). This value applies
for all evaluations.

Input range

-999 - 999 mV

Default value

-15 mV

Max. offset potential

Maximum permitted offset voltage (voltage at pH = 7.0). This value
applies for all evaluations.

Input range

-999 - 999 mV

Default value

15 mV

6.3.3.6.6

Data for specified electrode types

Tab: Configuration

Extras Templates Electrode types... [Proper-

ties]

Very good electrode / Good electrode / Usable electrode

The three predefined electrode types Standard, Gel and Non-aqueous
are always present in the table and be neither edited nor deleted. The fol-
lowing limit values are defined for the evaluation of the electrodes of
these electrode types:

Electrode type "Standard"

Parameters

Excellent

electrode

Good

electrode

Usable

electrode

Max. streaming potential
(mV)

2.5

3.0

4.0

Max. drift (mV/min)

2.0

2.5

3.0

Min. slope (%)

96.5

96.0

95.0

Max. slope (%)

101.0

102.0

103.0

Max. response time (s)

45

50

60

Min. offset potential (mV)

-15

-15

-15

Max. offset potential (mV)

15

15

15