5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

902

■■■■■■■■

tiamo 2.4

Titrando, 855
Input range

0.1 - 2,000.0 mV

Default value

60.0 mV

Selection

off

Titrino
Input range

1.00 - 2,000 mV

Default value

60.0 mV

Selection

off

Max. rate

The maximum dosing rate depends on the cylinder volume of the
exchange unit or dosing unit used. The following rule of thumb should be
used as a starting point for parameterization: Max. rate in mL/min = 5 •
Expected rate of reaction in mL/min.

Titrando, 855
Input range

0.01 - 166.00 mL/min

Default value

0.25 (50 µL/min), 0.75 (100 µL/min), 2.00
(500 µL/min) mL/min

Titrino
Input range

0.01 - 150.00 mL/min

Default value

0.25 (50 µL/min), 0.75 (100 µL/min), 2.00
(500 µL/min) mL/min

Min. rate

Rate at which dosing is carried out at the very beginning of the titration
and in the Control range at the end of the titration. The smaller the
selected minimum rate, the longer it takes until the control point is
reached for the first time. The following rule of thumb should be observed
for maintaining the Min. rate in order to achieve as constant (regular) a
dosing at the control point as possible: Min. rate in µL/min = Expected
rate of reaction in µL/min/ 10.

Input range

0.01 - 9,999.00 µL/min

Default value

10.00 (50 µL/min), 20.00 (100 µL/min), 40.00
(500 µL/min) µL/min